UWSpace is currently experiencing technical difficulties resulting from its recent migration to a new version of its software. These technical issues are not affecting the submission and browse features of the site. UWaterloo community members may continue submitting items to UWSpace. We apologize for the inconvenience, and are actively working to resolve these technical issues.
 

Bayesian Filtering of Surface EMG for Accurate Simultaneous and Proportional Prosthetic Control

Loading...
Thumbnail Image

Date

2016-12-01

Authors

Hofmann, David
Jiang, Ning
Vujaklija, Ivan
Farina, Dario

Journal Title

Journal ISSN

Volume Title

Publisher

Institute of Electrical and Electronics Engineers

Abstract

The amplitude of the surface EMG (sEMG) is commonly estimated by rectification or other nonlinear transformations, followed by smoothing (low-pass linear filtering). Although computationally efficient, this approach leads to an estimation accuracy with a limited theoretical signal-to-noise ratio (SNR). Since sEMG amplitude is one of the most relevant features for myoelectric control, its estimate has become one of the limiting factors for the performance of myoelectric control applications, such as powered prostheses. In this study, we present a recursive nonlinear estimator of sEMG amplitude based on Bayesian filtering. Furthermore, we validate the advantage of the proposed Bayesian filter over the conventional linear filters through an online simultaneous and proportional control (SPC) task, performed by eight able-bodied subjects and three below-elbow limb deficient subjects. The results demonstrated that the proposed Bayesian filter provides significantly more accurate SPC, particularly for the patients, when compared with conventional linear filters. This result presents a major step toward accurate prosthetic control for advanced multi-function prostheses.

Description

Keywords

Bayesian Filter, EMG Amplitude Estimation, Simultaneous And Proportional Control

LC Keywords

Citation