dc.descriptionThis is an open access article published under an ACS AuthorChoice License, which permits copying and redistribution of the article or any adaptations for non-commercial purposes.http://pubs.acs.org/page/policy/authorchoice_termsofuse.html. This document appeared in final form in Analytical Chemistry, copyright © American Chemical Society http://dx.doi.org/10.1021/acs.analchem.6b01802en
dc.description.abstractWe present the kinetics of calibrant release and analyte uptake between the sample and calibrant-loaded extraction phase, CL-EP, with a finite-element analysis (FEA) using COMSOL Multiphysics software package. Effect of finite and infinite sample volume conditions, as well as various sample environment parameters such as fluid flow velocity, temperature, and presence of a binding matrix component were investigated in detail with the model in relation to the performance of the calibration. The simulation results supported by experimental data demonstrate the suitability of the CL-EP method for analysis of samples with variation of the sample environment parameters. The calibrant-loaded approach can provide both total and free concentrations from a single experiment based on whether the partition coefficient (Kes) value being used is measured in a matrix-matched sample or in a matrix-free sample, respectively. Total concentrations can also be obtained by utilizing CL-EP in combination with external matrix-matched calibrations, which can be employed to automate the sampling process and provide corrections for variations in sample preparation, matrix effects, and detection processes. This approach is also suitable for very small volumes of sample, where addition of an internal standard in the sample is either troublesome or can change the sample characteristics.en
dc.description.sponsorshipNational Science and Engineering Research Council (NSERC)en
dc.description.sponsorshipOntario Research Fund (ORF) of Canadaen
dc.language.isoenen
dc.publisherAmerican Chemical Societyen
dc.rightsStandard ACS AuthorsChoice/Editors' Choice Usage Agreement*
dc.rights.urihttp://pubs.acs.org/page/policy/authorchoice_termsofuse.html*
dc.subjectHydrophobic Organic Contaminantsen
dc.subjectPolyethylene Passive Samplersen
dc.subjectMicroextraction Spmeen
dc.subjectKinetic Calibrationen
dc.subjectWateren
dc.subjectChemicalsen
dc.subjectTransporten
dc.subjectBindingen
dc.subjectFiberen
dc.subjectTimeen
dc.titleNumerical Simulation and Experimental Validation of Calibrant-Loaded Extraction Phase Standardization Approachen
dc.typeArticleen
dcterms.bibliographicCitationAlam, M. N., & Pawliszyn, J. (2016). Numerical Simulation and Experimental Validation of Calibrant-Loaded Extraction Phase Standardization Approach. Analytical Chemistry, 88(17), 8632–8639. https://doi.org/10.1021/acs.analchem.6b01802en
uws.contributor.affiliation1Faculty of Scienceen
uws.contributor.affiliation2Chemistryen
uws.typeOfResourceTexten
uws.peerReviewStatusRevieweden
uws.scholarLevelFacultyen


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record


UWSpace

University of Waterloo Library
200 University Avenue West
Waterloo, Ontario, Canada N2L 3G1
519 888 4883

All items in UWSpace are protected by copyright, with all rights reserved.

DSpace software

Service outages