dc.contributor.authorAlizadeh, Mostafa
dc.date.accessioned2019-01-23 16:37:08 (GMT)
dc.date.available2019-01-23 16:37:08 (GMT)
dc.date.issued2019-01-23
dc.date.submitted2019-01-21
dc.identifier.urihttp://hdl.handle.net/10012/14392
dc.description.abstractA vision on the migration from contact standard health monitoring measurement devices to non-contact measurement technologies has gained a tremendous attention in literature and in industry. A promising method for realizing the remote measurement of vital signs is using electromagnetic radars such as frequency modulated continuous wave (FMCW) radars. However, using these radars has challenges to precisely acquire the respiration and heart rates. A solution for higher accurate measurement of the vital signs can be the use of mm-wave frequencies, which gives a high-resolution sensing of displacements in an environment in the order of sub-mm changes. On the other hand, being in mm-wave bands increases both hardware and signal processing designs and implementations. In this work, a mm-wave radar is used to monitor the breathing and the heart rates as well as their waveforms for further clinical diagnostics. To that end, we established a complete analysis of the FMCW radars principles by considering hardware impairments. The analysis considers the effect of antenna coupling, RF cross-talk, stationary clutters, phase noise, IQ imbalances, and the thermal noise. Also, the effect of the individual hardware imperfections on the phase quality is shown by simulations and experiments. The simulations are carried out with a Matlab Simulink model. For the experiments, Texas Instruments (TI) mm-wave FMCW radars have been used. To earn insight into vital signs monitoring, different experiments are designed. In the experiments, the effect of the thermal instability of the RF parts on the phase is shown. In addition, to mimic the behaviour of the chest vibration due to respiration and the heartbeats, a two-pendulum system is designed and tested. Particularly, the pendulum system performance in terms of vibration frequency estimations of the two pendulums versus distance is then measured. In the simulations, the system performance is obtained for different signal to noise ratios (SNR) and different phase noise levels, as well as different stationary clutters. Finally, to test the TI sensors for different directions to the subjects, Hexoskin smart garment is used as a reference sensor, which is a reliable commercial product. Our results show great system improvement in terms of accuracy of the vital signs detection in comparison to other similar research. For different sleep positions, the accuracy of HR and BR are greater than 94\% and 96\%, respectively. In addition to detecting the vital rates, we have shown that their waveforms can also be reconstructed by using an adaptive optimum filter.en
dc.language.isoenen
dc.publisherUniversity of Waterlooen
dc.subjectfmcw radaren
dc.subjectmm-waveen
dc.subjectvital signsen
dc.subjectbreathing rateen
dc.subjectheart rateen
dc.subjectvibration detectionen
dc.subjectHMUSICen
dc.subjectharmonic analysisen
dc.subjectMatlaben
dc.subjectpendulumen
dc.subjectSimulinken
dc.titleRemote vital signs monitoring using a mm-wave FMCW radaren
dc.typeMaster Thesisen
dc.pendingfalse
uws-etd.degree.departmentElectrical and Computer Engineeringen
uws-etd.degree.disciplineElectrical and Computer Engineeringen
uws-etd.degree.grantorUniversity of Waterlooen
uws-etd.degreeMaster of Applied Scienceen
uws.contributor.advisorSafavi-Naeini, Safieddin
uws.contributor.advisorShaker, George
uws.contributor.affiliation1Faculty of Engineeringen
uws.published.cityWaterlooen
uws.published.countryCanadaen
uws.published.provinceOntarioen
uws.typeOfResourceTexten
uws.peerReviewStatusUnrevieweden
uws.scholarLevelGraduateen


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record


UWSpace

University of Waterloo Library
200 University Avenue West
Waterloo, Ontario, Canada N2L 3G1
519 888 4883

All items in UWSpace are protected by copyright, with all rights reserved.

DSpace software

Service outages