dc.description.abstractMany medical imaging modalities, such as mammography and micro-computed tomography, utilize digital X-ray imagers to observe human anatomy. Direct digital X-ray imagers rely on a sensor layer (typically a photoconductor) to convert X-ray photons to electrical charge, which can then be collected by pixel readout circuits. Whereas traditional integration-mode X-ray imagers typically integrate charge for long durations to acquire image frames, X-ray photon-counting imagers (PCI) resolve each incident photon as it arrives. This allows for equal energy-weighting of photons and multi-spectral image capture, both of which enhance contrast in images. Furthermore, PCIs also allow for higher dynamic range since count rates are not limited by integration well capacity. Many hybrid X-ray PCIs have been reported in the literature using photoconductors such as CdTe and HgI₂. However, these photoconductors are expensive to fabricate, suffer from low yield over large areas, and have limited spatial resolution. This thesis describes the design and characterization of the first hybrid X-ray amorphous selenium–CMOS PCI for mammography and micro-computed tomography. Amorphous selenium (a-Se) can be thermally deposited over large areas, allowing for cheaper and scalable fabrication as well as higher spatial resolution. Two arrays of 26 × 196 pixels are implemented in CMOS and interface directly to an a-Se sensor layer. Counter arrays neighbor the pixel arrays and have a one-to-one relationship with pixels, incrementing every time a photon is detected. Novel readout circuits allow for ultra high-resolution pixels, each occupying only 11.44 × 11.44 μm². Finally, the design of a custom PCB and FPGA system for characterizing the electronic performance of the PCI is described. The measured input-referred noise and threshold spread of the PCI are 41 e-rms and 107 e-rms, respectively, when operating the imager as a row scanner. This will enable an energy resolution of 5.7 keV, suitable for the proposed applications. Further analysis was done to identify methods of reducing threshold spread as well. Finally, the concluding chapter summarizes this work, compares its performance to other PCIs in the literature, and identifies future work to improve its performance.en
dc.language.isoenen
dc.publisherUniversity of Waterlooen
dc.subjectX-rayen
dc.subjectImagingen
dc.subjectCMOSen
dc.subjectseleniumen
dc.subjectphoton-countingen
dc.subjecthybrid imageren
dc.subjectmammographyen
dc.subjectcomputed tomographyen
dc.subjectdual-energyen
dc.subjectmultispectralen
dc.titleHybrid Amorphous Selenium-CMOS Photon-Counting X-ray Imageren
dc.typeMaster Thesisen
dc.pendingfalse
uws-etd.degree.departmentElectrical and Computer Engineeringen
uws-etd.degree.disciplineElectrical and Computer Engineeringen
uws-etd.degree.grantorUniversity of Waterlooen
uws-etd.degreeMaster of Applied Scienceen
uws-etd.embargo.terms1 yearen
uws.contributor.advisorKarim, Karim
uws.contributor.advisorLevine, Peter
uws.contributor.affiliation1Faculty of Engineeringen
uws.published.cityWaterlooen
uws.published.countryCanadaen
uws.published.provinceOntarioen
uws.typeOfResourceTexten
uws.peerReviewStatusUnrevieweden
uws.scholarLevelGraduateen


Files in this item

Thumbnail

This item appears in the following Collection(s)

Show simple item record


UWSpace

University of Waterloo Library
200 University Avenue West
Waterloo, Ontario, Canada N2L 3G1
519 888 4883

All items in UWSpace are protected by copyright, with all rights reserved.

DSpace software

Service outages